LDZ-5/S1  
 
click to enlarge
 
Test cell for solid dielectric materials

 
Applications:
  • Applicable in connection with the DSP-bsed Dissipation Factor & Capacitance Measuring System LDV-5 or the Dielectric Analyser DIANA
  • Test cells are designed according to the national and international standards DIN 53483 part 3 and VDE 0303 and IEC
  • Measuring the capacitance, the loss factor and the relative permittivity
 
LDZ-5/S1 Brochure
LDZ-5/S1 Technical Paper
LDZ-5/S1 Demo Software
LDZ-5/S1 Display Samples
LDZ-5/S1 RFQ
LDZ-5/S1 Slide Show
Get Adobe Acrobat Reader!  
 
 
Specifications:
  • Storage temperature - +15°C to +55°C
  • Storage humiditity - 90 %
  • Operation temperature - +5 °C to +40°C
  • Operation humidity - 85 %
  • Protection class - IP 00
  • Connections
    • BNC for the measuring electrode
    • Lab socket for the high voltage electrode
    • Lab socket for the guard electrode
  • Maximum test voltage (depend on the thickness of the test sample) - 2 kV AC r.m.s.
  • Dimensions (l x h x w) - approx. 170mm x 135 mm x 210 mm
  • Weight (total) - approx. 4.2 kg
  • Weight of the high voltage electrode alone - approx. 2.0 kg
  • Diameter measuring electrode - 100 mm
  • Width guard gap - 1mm
  • Width guard electrode - 20mm
  • Diameter high voltage electrode - 142
  • Electrode material - Stainless steel, polished
Return to top
 
Characteristics:
  • Consists of two parallel plate electrodes
  • Measuring electrode is shielded by a guard electrode for the elimination of stray capacitances
  • Fixed gauge is measuring the thickness of the insulating material
  • Additional possibility to measure the resistivity of the insulating material and the surface resistance in connection with a suitable measuring instrument, such as the Digital TERA-Ohmmeter DT-08
  • Due to the compact design the test cell can be placed inside the HV test box of the Dielectric Analyser DIANA