LDZ-5/LD  
 
click to enlarge
 
Test cell for liquid dielectric materials

 
Applications:
  • Applicable in connection with the DSP-based Dissipation Factor & Capacitance Measuring System LDV-5 or the Dielectric Analyser DIANA
  • Test cells are designed according to the national and international standards DIN 53483 part 3 and VDE 0303 and IEC
  • Measuring the capacitacne, the loss factor and the relative permittiviry (LDZ-5/LD especially for insulating oils)
 
LDZ-5/LD Brochure
LDZ-5/LD Technical Paper
LDZ-5/LD Demo Software
LDZ-5/LD Display Samples
LDZ-5/LD RFQ
LDZ-5/LD Slide Show
Get Adobe Acrobat Reader!  
 
 
Specifications:
  • Miximum test voltage - 2 kV AC
  • No-load capacitance - 50 pF
  • tan (air) - < 3 x 10 -5
  • Electrode distance - 2 mm
  • Electrode material - Stainless steel, preicision manufactured
  • Liquid quantity - approx. 30 ml (max. 150°C)
  • Operation temperature - +5°C to +60°C
  • Storage temperature - -15°C to +60°C
  • Accuracy of the thermometer - ±1°C
  • Connections -
    • LEMOSA-socket for the measuring voltage
    • LEMOSA-socket for the high voltage
    • Lab socket for additional grounding
  • Dimensions (diameter x H) - approx. 150 mm x 200 mm
  • Weight - approx. 6 kg
Return to top
 
Characteristics:
  • Consists of two coaxial stainless steel electrodes
  • Measuring electrode is shielded by a guard electrode for the elimination of stray capacitances
  • Temperatures of the measuring electrode and the high voltage electrode can be read by digital thermometers separately
  • Through the complete insulation of the live parts a safety operation is possible
  • Only little quantities of the liquid insulating material will be needed (30
Return to top