Double Pulse Generator LDC-5/D  
 
 
Double Pulse Generator LDC-5/D

 
Applications:
  • PD measurements on long cables and in conformance with standards require the verification of the double pulse behaviour of the employed PD measuring system
  • According to the IEC 855-3 the Double Pulse Generator LDC-5/D produces these standardized time parameters
  • In addition there is an output for calibrating pulse charges (5, 20, 100, 500 pC).
 
LDC-5/D Brochure
LDC-5/D Technical Paper
LDC-5/D Demo Software
LDC-5/D Display Samples
LDC-5/D RFQ
LDC-5/D Slide Show
Get Adobe Acrobat Reader!  
 
 
Specifications:
  • Pulse shape
    • Rise time (10 % .. 90 %) - < 20 ns
    • Pulse duration (10 % .. 10 %) - < 150 ns
    • Double pulse distance (10 % .. 10 %) - < 0.2 µs < t < 100 µs
  • Pulse charge injected into the impedance of 50 Ohm - approx. 100 pC
  • Pulse repitition rate - approx. 600 Hz
  • Calibrating charges - 5; 50; 100; 500 pC or 0,5; 2; 10; 50 pC
  • Duration of charge injection - < 50 ns
  • Powering voltage - 9 V (3.g. 6LR61)
  • Current consumption - 25 mA
Return to top
 
Characteristics:
  • Contains a pulse amplifier and a peak detector
  • Possibility to connect a scope to monitoring PD events (DC coupling and a sensitivity of 0.5 V/div is reccomended)
  • Powered by : 9 V battery (6LR61), current consumption less than 10 mA
  • BNC-Adapter, fibre optic cable 10m
Return to top